A scattering correction method for a panel detector based cone beam computed tomography system is presented. First,the x-ray spectrum of the system is acquired by using the Monte Carlo simulation method.Secondly,scattered photon distribution is calculated and stored as correction matrixes by using the Monte Carlo simulation method according to scanned objects and computed tomography system specialties.Thirdly,scattered photons are removed from projection data by correction matrixes.A comparison of reconstruction image between before and after scattering correction demonstrates that the scattering correction method is effective for the panel detector based cone beam computed tomography system.
First, the x-ray spectrum of the system is acquired by using the Monte Carlo simulation method. Secondarily, scattered photon distribution is calculated and stored as correction matrixes by using the Monte Carlo simulation method according to scanned objects and computed tomography system specialties. Thirdly, scattered photons are removed from projection data by correction matrixes. A comparison of reconstruction images between before and after scattering correction demonstrates that the scattering correction method is effective for the panel detector based cone beam computed tomography system.