热压灭菌时输液瓶炸裂的工伤事故,我区时有发生。为了探讨防范措施,现对事故原因探讨如下: 1 排汽管道不畅通 卧式高压灭菌柜的底部有排汽管,主要排出灭菌柜内的冷却水及少量蒸汽。该灭菌柜的温度感受器是装在底部排气管内,当管内存有冷却水时,其感受器的温度与灭菌柜内的
Sterilization infusion bottle burst injury accident, my area has occurred. In order to explore preventive measures, the cause of the accident are as follows: 1 exhaust pipe is not smooth Horizontal autoclave bottom of the exhaust pipe, the main discharge of sterilized cabinet cooling water and a small amount of steam. The sterilizer temperature sensor is installed in the bottom of the exhaust pipe, when there is cooling water pipe, the susceptor temperature and the sterilizer inside