故障现象 该机心内去颤电极板在使用一段时间后,会出现不能去颤现象。表现为,当主机充好电后,按下电极板手柄开关时,去颤电极无电压输出,主机屏幕上显示,去颤电压强度值慢慢下降到零。故障检修 检测电极板,插头1、2脚分别接两个去颤电极,其3、5脚接去颤开关,用万用表测3、5脚开关有较大的接触电阻,且阻值很不稳定,可以断定是开关损坏,叮能是电极板反复经过高温消毒引起开关触头氧化。心内去颤电极板是全封闭不可拆卸的,无法直接修理,我们在主机对应的插座3、5脚上加焊一个微动开关替代手柄上之开关,并将开关固定在机壳内侧上。使用时。由手术者指挥台下工作人员按此开关进行去颤放电,这样可继续使用原来的电极板,且更换新电极也不会受影响。
Symptoms After the defibrillation electrode plate in the machine is used for a period of time, there will be no defibrillation. The performance is that when the host computer is charged, when the electrode plate handle switch is pressed, the defibrillation electrode has no voltage output, and the main screen shows that the intensity of the defibrillation voltage gradually drops to zero. Fault detection electrode plate, plug 1 and 2 feet respectively connected to two defibrillation electrodes, the 3, 5 feet to the tremor switch, with a multimeter to measure the 3,5-pin switch has a larger contact resistance, and the resistance is very unstable , It can be concluded that the switch is damaged, and the reason is that the electrode plate is repeatedly sterilized by high temperature sterilization to cause oxidation of the switch contacts. The defibrillation electrode plate is completely closed and cannot be directly repaired. We add a microswitch to replace the handle on the sockets corresponding to the sockets 3 and 5 of the host, and fix the switch on the inner side of the case. when using it. The defibrillation discharge is performed by the switch staff under the operator’s podium, so that the original electrode plate can be continuously used and the replacement of the new electrode will not be affected.