论文部分内容阅读
本文报道1例经股动脉血管闭合器(Angio-Seal)破裂和栓塞的罕见病例,这一罕见情况导致该例确诊有外周血管疾病的糖尿病患者在经皮冠状动脉介入后出现了急性下肢缺血。股浅动脉栓子切除术显示聚合固定器破裂成两部分,并嵌入一个大的血栓(25mm×5mm)中。
This article reports a rare case of rupture and embolization of the femoral artery occluder (Angio-Seal), a rare case of diabetic patients with peripheral vascular disease diagnosed with peripheral vascular disease presenting with acute lower limb ischemia after percutaneous coronary intervention . Superficial femoral artery embolization showed the polymeric fixator ruptured into two parts and embedded in a large thrombus (25 mm × 5 mm).