手提式灭菌锅虽然具有结构简单、便于操作、携带轻便、经济实用等特点。但也会出现故障,原则上应该更换,但也可在不得已时酌情修理。现就常见的故障与排除方法简介于下。 1.锅盖密封不严,有漏气现象发生,该故障多因橡胶垫圈老化或表面不平所致。排除方法:①更换垫圈。先将老化的垫圈去除,将垫槽内粘着的橡胶残渣刮洗干净。取同规格的新垫圈装入垫槽,再用木制小锤轻轻敲平即可;②垫圈局部收缩表面四凸不平造成漏气现象发生。这种故障
Although the portable sterilizer has a simple structure, is easy to operate, portable, economical and practical. However, failures can occur. In principle, they should be replaced, but they can also be repaired when necessary. The common faults and troubleshooting methods are described below. 1. The lid is not tightly sealed and there is a phenomenon of air leakage. This malfunction is caused by the aging of the rubber gasket or the unevenness of the surface. Remedy: 1 Replace the gasket. Remove the aged gasket first, and scrape the rubber residue stuck in the pad tank. Take the new gasket with the same specifications into the pad groove, and then gently tapping with a wooden hammer; 2 gasket shrinking the surface of the four convex unevenness caused by leakage phenomenon. This failure