手提式热压灭菌器由于体型小,不受环境所限,适用于各种类型的炉子,在农村卫生室及基层医疗单位广为采用。然而,许多单位由于对压力装置及灭菌原理不理解,而忽视了小小放气软管的使用。很多灭菌器的放气软管锈断,还继续使用,未能及时修理或更换,这样做是省钱了,但既危险,又达不到消毒灭菌的效果。 笔者曾经做过二组实验:18批不同批号的检品,分成二组进行高压灭菌。第一组用正常的热压蒸气灭菌器,第二组用放气软管锈断重新接上的灭菌器。两组均采用120℃,15磅,30分钟的消毒条件。相同
Portable autoclave due to small size, free from the environment, suitable for all types of stoves, widely used in rural clinics and primary health care units. However, many units ignore the use of small deflation hoses due to the lack of understanding of pressure devices and sterilization principles. Many sterilizers deflated hose rust, but also continue to use, failed to repair or replace in time, this is to save money, but not only dangerous, but also failed to sterilize and sterilize. I have done two sets of experiments: 18 batches of different batches of seized goods, divided into two groups autoclaved. The first group uses a normal autoclave and the second uses a deflation hose to reconnect the sterilizer. Both groups used 120 ° C, 15 pounds, 30 minutes of disinfection conditions. the same