论文部分内容阅读
本文报告了输液泵在心功能衰竭患者用硝普钠纠正心衰时的应用,通过分组观察输液速度及不良反应,证明应用输液泵输液,速度的符合率为100%,没有严重的不良反应,只有恶心、呕吐,不良反应发生率为1.4%。而普通静脉输液速度符合率为56%,不良反应为眩晕、大汗为4%,恶心、呕吐为8%。结果证明,输液泵在临床上能保证输液速度的准确性,保证用药质量,降低药物的不良反应。近十几年来,我们应用硝普钠纠正心功能衰竭进展很快、疗效很好。但该药对输液速度要求严格,速度稍过快则发生明显不良反应。近几年,我们通过应用微量输液泵输
This article reports the application of infusion pump in patients with heart failure with sodium nitroprusside to correct heart failure. By grouping the infusion rate and adverse reactions, it was proved that the rate of compliance with infusion pump and infusion rate was 100% and no serious side effects Nausea and vomiting, the incidence of adverse reactions was 1.4%. The coincidence rate of ordinary intravenous infusion rate was 56%, adverse reactions were dizziness, sweating 4%, nausea, vomiting 8%. The results show that the infusion pump in clinical practice to ensure the accuracy of infusion speed, to ensure the quality of medication, reduce adverse drug reactions. In the past ten years, we applied sodium nitroprusside to correct heart failure and achieved rapid success. However, the drug on the infusion rate of strict requirements, slightly faster rate of significant adverse reactions. In recent years, we use the application of micro infusion pump