F-800自动血球计数仪故障分析检修

来源 :中国医疗器械杂志 | 被引量 : 0次 | 上传用户:cx77287728cx
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
故障一:WASTEFULL(废水瓶满)出现该故障报警时,一般是废水瓶内废液已满,倒掉即可。如果废液未满,则是水位传感器或其相关电路故障。这时可把水位传感器的插头从仪器的后面拔下,然后短接该插座。如果不报警了,则是传感器或其引线开路。由于使用过程中经常要开盖倒废液, Fault 1: WASTEFULL (drink bottle full) When this fault alarm occurs, the waste liquid in the waste bottle is generally full, and it can be drained. If the waste is not full, the water level sensor or its related circuit is faulty. At this time, you can unplug the water level sensor from the back of the instrument and short-circuit the socket. If no alarm is given, the sensor or its lead is open. Due to the frequent use of liquid waste during use,
其他文献
请下载后查看,本文暂不支持在线获取查看简介。 Please download to view, this article does not support online access to view profile.
期刊
故障现象:机器开机后,power指示灯亮,但无声,显示屏无图。分析检修过程:从故障现象可以初步判断此故障出在高压振荡电路部分。打开机器盖板,取下高压电路板(13-1—OO24)及高压屏蔽盒(参
早产儿因发育不全,易患颅内出血,本文介绍一种无损检测的仪器,用可见光从新生儿前隆突投照,从囟门提出光信号,经单片机运算、判断,迅速查出颅内是否有出血,为治疗赢得宝贵时间。 Pr
四月一日至八日,本刊编辑部邀集了我省部分专业、业余词曲作者,集中时间精力,进行歌曲创作和讨论,以期创作出有较高水平或较好基础的歌曲作品。这次会议,是按照省音协二届四
请下载后查看,本文暂不支持在线获取查看简介。 Please download to view, this article does not support online access to view profile.
期刊
目的:探讨内皮素-1(ET-1)在喉癌组织中的表达及其意义。方法:82例喉癌肿瘤组织(TNM分期:Ⅰ期27例,Ⅱ期34例,Ⅲ期21例)和82例喉正常黏膜组织、80例喉良性病变组织为研究对象,
美国Baxter公司生产的SPS-550型血透机是采用微机处理,具有自动超滤控制(UFC)功能,可做醋酸和碳酸二种病人的机器。它的浓度配比系统采用比例,这部分故障相对较多,工作时受其它方面
医疗设备是医院进行诊疗工作,开展科学研究和临床教学的物质条件,是医院现代化建设的一个重要方面。在目前国防经费紧缩、医院财力不足的情况下,采取“两条腿”走路方针,缓解了医
我院于93年引进一台美国GANDCOR动态心电监护仪,最近连续两次发生QRS波形异常故障。现将这两次故障排除经验介绍如下:故障一:装好回放磁带,按常规顺序开机进入运行程序后,屏幕下方