日立EUB—25M型B超仪多重故障检修一例

来源 :医疗装备 | 被引量 : 0次 | 上传用户:xionglongyan0817
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
故障现象:日立EUB—25M型B超仪,因电子交流稳压器故障,观察、照相监示器突然无光。冻结指示灯不亮,冻结开关也失灵,但是散热风扇工作,各控制指示灯亮。分析检修:很明显由于两个监视器同时无光,应首先检查电源。开盖查保险丝F114、F113(15A)均熔断 Symptoms: Hitachi EUB-25M B-type ultrasonic instrument, due to the failure of the electronic AC voltage regulator, the observation and camera monitor suddenly become dull. The freeze indicator does not light, and the freeze switch also malfunctions, but the cooling fan works and each control indicator lights up. Analysis and Maintenance: It is obvious that since the two monitors are not at the same time, the power supply should be checked first. Open cap check F114, F113 (15A) are all blown