论文部分内容阅读
预真空压力蒸气灭菌器对标准新旧手术布巾B-D试验包测试各54次,并与3M化学指示胶带自制指示图对照。测试结果表明物品包巾B-D试验成功与否,与使用新旧布巾有关系。B-D试验化学指示图与3M化学指示胶带自制指示图,对物品包中空气团敏感性试验有显蓍性差异,3M胶带自制图不能有效地反映灭菌器内空气气团的存在。
Pre-vacuum pressure steam sterilizer on the standard old and new surgical cloth towel B-D test package test 54 times each, and with the 3M chemical indicator tape homemade indication control. The test results show that the success of the package B-D test towel bag, and the use of old and new cloth towels have a relationship. The B-D test chemical indicator and the 3M chemical indicating tape self-made indicator showed obvious difference in the air mass sensitivity test of the bag. The 3M tape self-made map can not effectively reflect the presence of air mass in the sterilizer.