论文部分内容阅读
我院于1996年7月购进美国柯达90型自动洗片机,根据我院一年多的使用、观察和保养维修,对该型机易出现的故障,特做如下介绍:1.异常声响 洗片过程中,有时会出现异常声响,原因。①因齿轮干燥,摩擦增大,致使辊轴齿轮与驱动轴齿轮咬合错位。②进片时胶片摆放不正,发生卡片。③药液因使用时间长,传送过程中,定影液外渗,久之使得药液结晶析出沉积在齿轮上,使齿轮运转不平衡发生异常咔咔声音。 处理:①缩短药液使用周期,降低药液粘稠度。②定期冲洗齿轮上的结晶,并加润滑油,使运转平稳。③如遇卡片,打开干燥箱取出卡住的胶片,输送胶片时仔细认真。
In July 1996, our institute purchased the Kodak Model 90 Automatic Tablet Processor. According to the use, observation, maintenance and repair of our hospital for more than one year, the failure of this type of machine is described as follows: 1. Abnormal noise In the process of washing, abnormal noise may sometimes occur. 1 Due to gear drying, the friction increases, resulting in the misalignment of the roller shaft gear and the drive shaft gear. 2 When the film is loaded, the film is not placed properly and the card is generated. (3) Due to the long time of use of the liquid medicine, during the process of transfer, the fixing liquid is extravasated, and the liquid crystal of the liquid is precipitated and deposited on the gear for a long time, so that abnormal gear noise may occur due to unbalanced operation of the gear. Treatment: 1 shorten the use of liquid medicine, reduce the liquid viscosity. 2 Regularly flush the crystals on the gears and add lubricating oil to make the operation smooth. 3 In case of a card, open the oven to remove the jammed film and carefully transport the film.