论文部分内容阅读
目的评价该实验室目前的25个常规生化检验项目参考区间是否适用于所服务的人群,验证参考区间的适用性。方法依据《WS/T 402-2012临床实验室检验项目参考区间的制定》的规定,每个项目至少选择20例健康参考个体,在该实验室3个分析系统(Beckman公司DxC800、AU5421型以及雅培公司c16000型)上,进行了25个常规生化检验项目的参考区间验证。结果该实验室所开展的25个常规生化检验项目的参考区间均通过验证。结论该实验室现在使用的25个常规生化检验项目参考区间适用于所服务的人群。
Objective To evaluate whether the reference interval of 25 routine biochemical test items in the laboratory is suitable for the population served and to verify the applicability of the reference interval. Methods According to the provisions of “WS / T 402-2012 clinical laboratory test project reference interval,” each project selected at least 20 healthy reference individuals in the laboratory three analysis system (Beckman DxC800, AU5421 type and Abbott Company c16000 type), the 25 routine biochemical test project reference interval validation. Results The reference intervals of 25 routine biochemical test items carried out by this laboratory were verified. Conclusion The 25 routine biochemical test project reference intervals currently used in this laboratory are appropriate for the population being served.