目的研究ATP生物荧光法用于气管镜清洗质量及消毒效果评价的有效性。方法选取临床检查中已使用过的气管镜共30件,每件均在清洗前、消毒前和消毒后3个时段,采集气管镜管腔冲洗水样品各1份,分别进行ATP生物荧光法检测;并对消毒后采样的冲洗水样进行菌落计数检测。结果清洗后,管腔冲洗水中的ATP含量明显下降,中位数从7 503 RLU降到42 RLU(P<0.05),清洗合格率为93.33%;消毒后ATP值中位数为14 RLU,均在合格范围内,但有消毒后ATP值高于消毒前的情况,且与菌落计数法结果有出入(合格率96.67%)。结论 ATP生物荧光法适用于评价器械清洗效果,检测结果能反映实际清洗质量,但不适用于器械消毒效果的评价。
Objective To study the effectiveness of ATP biofluorescence in assessing the quality of bronchoscopy and disinfection. Methods A total of 30 bronchoscopes used in clinical examination were selected, each of which was collected before the cleaning, before disinfection and after the disinfection for 3 periods, and one piece of bronchoalveolar lavage water sample was collected for ATP biofluorescence assay ; And after disinfection sampling flushing water samples colony count detection. Results After washing, the content of ATP in the lumen flushing water decreased obviously, the median decreased from 7,503 RLU to 42 RLU (P <0.05), and the pass rate of cleaning was 93.33%. The median ATP value after disinfection was 14 RLU In the qualified range, but after disinfection ATP value higher than before disinfection, and the colony count method results (pass rate 96.67%). Conclusion ATP biofluorescence method is suitable for evaluating the cleaning effect of instruments. The test results can reflect the actual cleaning quality, but it is not suitable for the evaluation of the disinfection effect of instruments.